Metrohm Power Box (for 899 Coulometer) User Manual
Page 16

4.2 Operating modes
■■■■■■■■■■■■■■■■■■■■■■
10
■■■■■■■■
Power Box
The Power Box is charged no matter whether the 899 Coulometer is
switched on or switched off.
899 Coulometer with the Power Box without power failure
mode
■
The power supply unit is connected to the Power Box and to the mains
supply.
■
The Power Box is switched off.
■
The 899 Coulometer is connected to the Power Box.
■
The 899 Coulometer is switched on.
The 899 Coulometer can be operated normally as long as the mains sup-
ply is stable. The Power Box does not take over supplying power if the
mains supply fails. The 899 Coulometer is shut down in a controlled way.
899 Coulometer with the Power Box without power supply
unit
■
The Power Box is fully charged and switched on.
■
The Power Box is not connected to the power supply unit and the
mains supply.
■
The 899 Coulometer is connected to the Power Box.
■
The 899 Coulometer is switched on.
The 899 Coulometer can be operated for at least 12 h (see Chapter 4.4,
page 12).
Stand-alone operation
■
The power supply unit is connected to the Power Box and to the mains
supply.
■
The Power Box is switched on.
The Power Box is charged. Once it is fully charged, the LED switches to a
higher light intensity (1-1).
The intelligence of the Power Box prevents overcharging, even if the
Power Box is connected to the mains supply for more than 14 h.