3.1 Attaching the bottle cap
■■■■■■■■■■■■■■■■■■■■■■
4
■■■■■■■■
IC equiment: Dosino Regeneration
3 Installation
Proceed as follows to install the IC equiment: Dosino Regeneration:
■
Attach the bottle cap to the regeneration solution bottle.
■
Assemble the 800 Dosino and fasten it to the holder.
■
Connect the Dosino to the bottle.
■
Connect the Dosino to the inline filter.
■
Connect the inline filter to the suppressor.
3.1
Attaching the bottle cap
Required accessories
■
Bottle with regeneration solution
■
Eluent bottle cap (6.1602.160)
■
Adsorber tube (6.1619.000)
■
Adapter for adsorber tube (6.1624.000)
■
FEP aspiration tubing (6.1829.020)
1 Mounting the aspiration tubing
■
Use the capillary cutter to cut the aspiration tubing to such a
length that it touches the bottom of the bottle.
■
Place the aspiration tubing in the M6 opening of the bottle cap.
■
Place an O-ring in the M6 opening.
2 Inserting the stopper
■
Tighten the M8 stopper to the M8 opening of the bottle cap.