2 detailed status conditions of the global $g, 3 detailed status conditions of the global $s, Detailed status conditions of the global $g – Metrohm 781 pH/Ion Meter: Operation via RS-232 User Manual
Page 7: Detailed status conditions of the global $s
1 General rules
780/781 pH/Ion Meter - Operation via RS232
5
1.3.2 Detailed
status
conditions of the global $G
$G.Mode.pH .Stirrer
: Waiting for stirring time during measurement.
.Cal .Inac
: At beginning of calibration.
.Stirrer
: Calibrating, waiting for stirring times.
.Req.Temp1
: Requesting temperature.
.Meas.TempX
: Temperature measurement in buffer X.
.Meas.BufX
: Calibrating, measuring buffer X.
.Req.BufX
: Requesting pH or buffer X.
.Data
: Display of data, data output.
.ElTest .Inac
: At beginning of electrode test.
.Req.TempX
: Requesting X
th
temperature.
.Meas.BufX
: Performing X
th
measurement.
.Req.BufX
: Waiting for X
th
buffer.
.Meas.IncX
: Measuring X
th
increment.
.Data
: Data output.
$G.Mode.T.Stirrer
: Waiting for stirring time during measurement.
$G.Mode.U.Stirrer
: Waiting for stirring time during measurement.
(781 only:)
$G.Mode.Conc .Add .Inac
: At beginning of standard addition.
.Stirrer
: Calibrating, waiting for stirring times.
.Req.Temp0
: Requesting temperature.
.Meas.Temp0
: Measuring temperature.
.Meas.Inc0
: Measuring initial solution.
.Req.IncX
: Requesting X
th
increment.
.Add.IncX
: X
th
increment is being added.
.Data
: Data output.
.Direct
.Stirrer
: Waiting for stirring time during measurement.
.Cal
.Inac
: At beginning of calibration.
.Stirrer
: Calibrating, waiting for stirring times.
.Req.Temp1
: Requesting temperature during calibration.
.Meas.Temp1 : Measuring temperature during calibration.
.Meas.StdX : Measuring X
th
standard.
.Req.StdX : Requesting X
th
standard.
.Add.StdX : X
th
standard is being added.
.Data
: Data output.
1.3.3 Detailed
status
conditions of the global $S
The process has been stopped arbitrarily because of an error. The condition from which the
process has been stopped is indicated. Example:
$S.Mode.Conc.Add.Inac
: Stopped at very beginning of standard addition.
.Stirrer : Stopped when waiting for stirring times.
$S.Mode.XX.Drift(Ok)
: Measurement stopped via RS232.