beautypg.com

Calibration, Specifi cations, Operation – Bacharach 50 User Manual

Page 2: Checking for co, Backlight, Overrange, Low battery, Ppm co, Snifit, Turning the snifi t on & off

background image

- 2 -

Figure 1. Snifi t 50

- 3 -

Figure 2. Calibration Equipment Setup

- 4 -

If you’re not sure about the quality of the surrounding

air, you can apply a blend of Oxygen/Nitrogen gas to

the sensor as described under Calibration.

If rapid temperature changes are encountered (such

as from freezing to room temperatures) during the

operation of the analyzer, the operator must allow the

analyzer to stabilize at the ambient temperature for at

least 2 minutes before taking a measurement.

Checking for CO

Important! Ensure that the sensor grille at the rear of

the instrument is unobstructed and open to the atmo-

sphere. A quick instrument check can be performed by

allowing the smoke of a blown-out match to enter into

the sensor grille. This should cause the indicated CO

level to increase.

After zeroing the Snifi t, simply hold the instrument in

the area that you suspect the presence of CO gas. If CO

is present, the instrument will indicate the concentrA-

tion of CO in ppm on its display.

For more extensive CO investigations and testing in

areas of higher temperatures, the Monoxor

®

II is the

preferred instrument.

Backlight

Pressing the Backlight button will illuminate the

display for a period of 8 minutes, or until it is manually

turned off by again pressing the Backlight button.

Overrange

If the CO level exceeds 1999 ppm, the

displayed reading is replaced by the

number “1”.

Low Battery

When the Snifi t’s 9V battery is nearing

the end of its useful life,

LO BAT

will

Calibration

Important! For accurate Snifi t operation, calibration

every 6 months is recommended.

To calibrate the CO sensor, you will need the equip-

ment listed under Accessories. Note that the Zero Gas

Cylinder is needed only if you’re unsure about the qual-

ity of the surrounding air for zeroing purpose.

Calibrate the sensor to a known concentration of CO

gas as follows:

1. Assemble the calibration equipment per Figure 2.

2. Zero the instrument as previously described under

Zeroing the Sensor.

If necessary, you can use the calibration cup to apply

a blend of Oxygen/Nitrogen gas directly over the sen-

sor by attaching a zero gas cylinder to the regulator

and adjusting the regulator knob for a fl ow rate of 2

SCFH.

MONOXIDE ANALYZER

MONOXIDE ANALYZER

ppm CO

CARBON

CARBON

Snifit

model 50

model 50

Backlight

On

Off

Span Adjust

-Power-

ZERO

ADJUST

(Recessed)

SPAN
ADJUST
POT.

BACKLIGHT

ON/OFF

POWER ON

POWER OFF

DISPLAY

LO BAT

LOW

BATTERY

INDICATION

1

2

3

4

Parts Shown:

1. CO Span Gas Cylinder, 51-1994
2. Regulator, 03-4318*
3. Tubing, 03-6351*
4. Flowmeter, 06-6163*
5. Calibration Cup Assembly, 23-2156**

* Part of Calibration Kit 24-7059
** Part of Tool / Cup Kit 19-3242

5

PUSH CALIBRATION CUP
OVER SENSOR HOUSING

LO BAT

Specifi cations

Gas Monitored .............Carbon Monoxide (CO)

Range ...........................0 to 1999 ppm

Resolution ....................1 ppm

Accuracy ......................± (5% of reading + 5 ppm)

Response ......................

30 sec. to 90% of fi nal value

Power ...........................9V battery

Battery Life .................1500 hours minimum

Temperature Range ....32 to 104 °F (0 to 40 °C)

Case Material ..............High impact plastic

Weight ..........................0.25 lb (0.11 kg) without battery

Dimensions ..................5.5" L x 2.0" W x 1.6" H

(140 x 51 x 41 mm)

Operation

Turning the Snifi t
On & Off

The Snifi t is turned on

and off by pressing the
Power On

and Power

Off

buttons, respec-

tively. If the instru-

ment is not manually

turned off, it will au-

tomatically shut itself

off after approximately

35 minutes.

Zeroing the
Sensor

Turn on the Snifi t and allow the displayed reading to

stabilize before proceeding – approximately 30 seconds.

Then with the instrument sampling fresh air (air that

is free of CO), use a screwdriver to adjust the Zero
Adjust

potentiometer at the top of the instrument for a

reading of 0 ±1 ppm.

appear in the upper left hand corner of the display.

Although you can continue using the instrument under

this condition, you should replace the battery as soon

as possible to ensure accurate CO readings.