SciGene 777 Microarray Oven User Manual
Page 12

Model 777 Microarray Oven — USER MANUAL
Version 2.1, February 2011 (408) 733-7337, [email protected]
11
V. TROUBLESHOOTING
Symptom
Cause
Solution
Rotator does not turn
after closing door
Rotator control is in the off
position
Turn on rotator
Door does not shut or
is difficult to open
Latch is improperly
adjusted
Adjust threaded
rod on inside of
latch
Temperature is erratic
Thermal controller is
defective
Replace controller
VI. SPECIFICATIONS
Electrical (Oven Unit)
Cat. #1077-00-1
115V AC; 50/60 Hz; 350 W
Cat. #1077-00-2
230V AC; 50/60 Hz; 350 W
Dimensions
Interior Chamber
14 W x 14 D x 14 H inches
36 W x 36 D x 36 H cm
Exterior
18 W x 17 D x 22 H inches
45 W x 40 D x 56 H cm
Weight
Net
70 lbs (32kg)
Gross
75 lbs (34kg) Includes shipping carton
Performance and Controls
Temperature Range
Ambient + 5°C to 99°C
Temperature Control
± 0.1°C
Heat up Time
3°C per minute
Temperature Controller
Digital PID, single loop
Temperature Display
Actual / Set Dual LED
Rotation Speed
Up to 80 RPM
Chamber
Stainless Steel
Digital Thermometer Output Thermocouple
VII. ROTATORS AND ACCESSORIES
Cat. #
Description
UOM
1070-20-0
Rotator for 20 Agilent SureHyb and Original Version
Hybridization Chambers.
Each
1077-10-0
Rotator for 48 Affymetrix GeneChips.
Each
1077-30-0
Rotator for 96 Affymetrix GeneChips.
Each
1051-52-0
Digital thermometer with NIST certificate of calibration
Each