Down loading data from the gx- 2001 – RKI Instruments GX-2001 User Manual
Page 13
Data Logging Software User’s Guide
Down Loading Data from the • 13
Down Loading Data from the
GX- 2001
You have the option of down loading data manually or automatically.
If you down load information using the automatic feature, click the
checkbox, “Automatic Download,” in the Download Window before placing
the Model GX-2001 in the Data Logging Station. Remember that the
Instrument Information screen will remain blank after Automatic
Downloading is selected.
NOTE: Before you attempt to down load data, make sure you have
installed the GX-2001 Data Logging Software and have connected
the serial data cable between the Data Logging Station and the
computer.
1. Launch the GX-2001 Data Logging Software. The Download Window
displays.
2. Place the GX-2001 into the Data Logging Station.
3. Choose the COM port on the computer to which you have connected
the Data Logging Station.
NOTE: The software includes a Cal.Unit Calibration button between the
Clear Data button and the Power Off button which is not shown in
this manual. This button is for use with the RKI Automatic
Calibration Station which is under development. If this button is
accidentally clicked, click the Cancel button which appears in
Automatic Calibration Mode screen.
Figure 8. Automatic Download Checkbox
Click here for
automatic
downloads