Start up, Operation, Maintenance/calibration – RKI Instruments GD-K71D4X User Manual
Page 8: Parts list

5 • Model GD-K71D4X Sample Draw Transmitter
CAUTION:
Leave the cable shield’s drain wire insulated and disconnected at the
GD-K71D4X. You will connect the opposite end of the cable shield’s drain wire at the
controller.
CAUTION:
At the controller, do not route power and GD-K71D4X wiring through the same
conduit hub. The power cable may disrupt the transmission of the GD-K71D’s
signal to the controller.
Start Up
See the GD-K71D Operator’s Manual for GD-K71D startup instructions.
Operation
See the GD-K71D Operator’s Manual for an operational description of the GD-K71D that
is mounted inside the enclosure.
Maintenance/Calibration
See the GD-K71D Operator’s Manual for maintenance and calibration instructions. Use
the inlet fitting on the NEMA 4X enclosure to apply gas instead of the fittings inside the
enclosure on the GD-K71D.
Parts List
Table 4 lists replacement parts and accessories for the GD-K71D.
Table 2: Parts List
Part Number
Description
06-1273RK
1/4” O.D. x 1/8” I.D. Teflon PTFE tubing
17-4820RK
Sample fitting, 1/4” tube bulkhead union
18-0107RK
3/4” conduit hub
33-0165RK-01
Hydrophobic filter w/ flexible tubing stubs