Troubleshooting – RKI Instruments 65-2504RK User Manual
Page 13

65-2504RK Oxygen Transmitter • 9
Troubleshooting
The troubleshooting guide describes symptoms, probable causes, and recommended
action for problems you may encounter with the oxygen transmitter.
NOTE:
This troubleshooting guide describes transmitter problems only. See the
controller instruction manual for problems you may encounter with the
controller.
Table 2:Troubleshooting the Oxygen Transmitter
Condition
Symptom(s)
Probable Causes
Recommended Action
Fail Condition
• Controller indicates a
fail condition.
• The transmitter wiring
is disconnected or
misconnected.
• The transmitter is
malfunctioning.
1. Verify that the transmitter wiring is
correct and secure.
2. Calibrate the transmitter.
3. If the fail condition continues, replace
the oxygen sensor.
4. If the fail condition continues, contact
RKI for further instruction.
Slow or No
Response/
Difficult or
Unable to
Calibrate
• Transmitter responds
slowly or does not
respond to response
test.
• Unable to accurately
set the fresh air or
zero reading during
calibration.
• Transmitter requires
frequent calibration.
Note: Under “normal”
circumstances, the
transmitter requires
calibration once every
three months.
Some applications may
require a more frequent
calibration schedule.
• The calibration cylinder
is low, out-dated, or
defective.
• The transmitter is
malfunctioning.
1. Verify that the calibration cylinder
contains an adequate supply of a
fresh test sample.
2. If the calibration/response difficulties
continue, replace the oxygen sensor.
3. If the calibration/response difficulties
continue, contact RKI for further
instruction.