Troubleshooting – RKI Instruments 65-2427RKSS-05 User Manual
Page 14

10 • 65-2427RKSS-05 H
2
S Detector
5.
Unscrew the calibration cup from the H
2
S detector.
6.
When the controller display reading falls below the alarm setpoints, return the
controller to normal operation.
7.
Store the components of the calibration kit in a safe place.
Quarterly
Calibrate the H
2
S detector as described in the Calibration section of this manual.
Troubleshooting
The troubleshooting guide describes symptoms, probable causes, and recommended
action for problems you may encounter with the H
2
S detector.
NOTE:
This troubleshooting guide describes detector problems only. See the controller
operator’s manual for problems you may encounter with the controller.
Table 2: Troubleshooting the H
2
S Detector
Condition
Symptom(s)
Probable Causes
Recommended Action
Fail Condition
Controller indicates a
fail condition.
• The detector wiring is
disconnected or
misconnected.
• The plug-in sensor is
not properly plugged
into the sockets in the
detector housing body.
• The detector’s zero
reading is low enough
to cause a fail
condition.
• The detector is
malfunctioning.
1. Verify that the detector wiring is
correct and secure.
2. Confirm that the plug-in sensor is
installed properly.
3. Perform a zero adjustment. A full
calibration is recommended.
4. If the fail condition continues, replace
the plug-in sensor as described later
in this section.
5. If the fail condition continues, contact
RKI for further instruction.
Slow or No
Response/
Difficult or
Unable to
Calibrate
• Detector responds
slowly or does not
respond during
response test.
• Unable to accurately
set the zero or
response reading
during calibration.
• Detector requires
frequent calibration.
Note: Under “normal”
circumstances, the
detector requires
calibration once every 3
months.
Some applications
may require a more
frequent calibration
schedule.
• The calibration cylinder
is low, out-dated, or
defective.
• The flame arrestor in
the detector housing
cap is wet or clogged
with dirt or other
particulates.
• The calibration gas is
not an appropriate
concentration.
• The detector is
malfunctioning.
1. Verify that the calibration cylinder
contains an adequate supply of a
fresh test sample.
2. Check the detector housing cap to
determine if the flame arrestor is wet
or dirty. Clean if necessary.
3. Verify that the calibration gas
concentration is appropriate for the
detector. Zero emission air (20.9%
oxygen) is normally used for a zero
adjustment if the environment is
suspect and 25 PPM H
2
S in nitrogen
is normally used for a response
adjustment.
4. If the calibration/response difficulties
continue, replace the plug-in sensor
as described later in this section.
5. If the calibration/response difficulties
continue, contact RKI for further
instruction.