Calibration frequency – RKI Instruments 65-2397RK User Manual
Page 19
65-2397RK CO
2
Transmitter • 15
NOTE:
When a transmitter is first powered up with a new amplifier, the initial output
may be either high or below zero depending on the setting of the zero pot. Be
sure to make arrangements so that this does not cause unwanted alarms.
10. Turn on power to the controller.
11. Turn on the controller and place it into normal operation.
NOTE:
When first powered up, the transmitter will enter about a one minute period
when the 4-20 mA output is stabilizing and may be above the controller alarm
points or well below zero momentarily. RKI controllers have a one minute
warmup period when the controller does not display any gas reading or give any
alarm indication. The CO
2
transmitter’s 4-20 mA signal should be stable by the
time the controller’s warmup period is over.
12. Allow the transmitter to warm-up for 5 minutes.
13. Calibrate the transmitter as described in “Calibration” on page 16 of this manual.
Calibration Frequency
Although there is no particular calibration frequency that is correct for all applications, a
calibration frequency of every 6 months is adequate for most infrared CO
2
transmitter
applications. Unless experience in a particular application dictates otherwise, RKI
Instruments, Inc. recommends a calibration frequency of every 6 months.
If an application is not very demanding, for example detection in a clean, temperature
controlled environment where calibration adjustments are minimal at calibration, then a
calibration frequency of every 9 to 12 months is adequate.
If an application is very demanding, for example if the environment is not well controlled,
then more frequent calibration than every 6 months may be necessary.
DETECTOR “W”
WHT
DETECTOR “G”
GREEN
DETECTOR “B”
BLK
Table 5:Reconnecting the CO
2
Detector to the Amplifier
Amplifier Detector
Terminal Strip
Detector Lead