Troubleshooting – RKI Instruments 61-0198RK User Manual
Page 13

61-1007RK/61-0198RK Carbon Dioxide Detector • 9
NOTE:
If the reading is not within ± 20% of the gas concentration, calibrate the detector
as described in “Calibration” on page 12.
9.
Turn the on/off knob clockwise to close the regulator.
10. Unscrew the regulator from the calibration cylinder and remove the sample tubing
from the regulator.
11. Unscrew the calibration cup from the detector. You may leave the sample tubing
connected to the calibration cup for convenience.
12. When the display reading falls below the alarm setpoints, return the controller to
normal operation or enable external alarms.
13. Store the components of the calibration kit in a safe place.
Biannually
Calibrate the CO
2
detector every six months as described in “Calibration” on page 12.
Troubleshooting
The troubleshooting guide describes symptoms, probable causes, and recommended
action for problems you may encounter with the CO
2
detector.
NOTE:
This troubleshooting guide describes detector problems only. See the controller
operator’s manual for problems you may encounter with the controller.
Table 2:Troubleshooting the CO
2
Detector
Condition
Symptom(s)
Probable Causes
Recommended Action
Fail Condition
• Controller indicates a
fail condition.
• The detector wiring is
disconnected or
misconnected.
• The plug-in sensor is
not properly plugged
into the three-socket
pattern in the detector
housing body.
• The detector zero
signal is low enough to
cause a fail condition.
• The detector is
malfunctioning.
1. Verify that the detector wiring is
correct and secure.
2. Confirm that the plug-in sensor is
installed properly.
3. Calibrate the detector.
4. If the fail condition continues, replace
the detector.
5. If the fail condition continues, contact
RKI for further instruction.