Quick reference guide for model eagle 2 – RKI Instruments Eagle 2 Quik Start User Manual
Page 2

Quick Reference Guide For Model Eagle 2
PN 71-0186RK Rev P1
RKI Instruments, Inc. 33248 Central Ave, Union City, CA 94587
Phone (510) 441-5656 • (800) 754-5165 • www.rkiinstruments.com
100430/1000
a. Perform a demand zero.
b. Cup hand over end of probe and gently exhale into probe.
c. In a few seconds, the oxygen reading will drop below the alarm
point of 19.5% and activate alarms.
Note: When testing using a hose, allow 1 second of time for each foot of hose
length. For example, allow a 5 foot hose 5 seconds for the sample to reach
the sensors.
d. Press POWER ENTER RESET to reset alarms.
6. Display Mode
a. To access Display Mode, while the Eagle 2 is in normal
operation, press and release DISPLAY ADJUST NO.
b. Use DISPLAY ADJUST NO to scroll through the following
screens: peak readings, battery voltage, gas display, methane
elimination mode (if unit is configured appropriately), relative
response (if RELATIVE RESPONSE is turned on), STEL,
TWA, view alarm settings, select user ID (if turned on), select
station ID (if turned on), time in operation, date/time, and data
logging time remaining (if turned on).
c. Press and release DISPLAY ADJUST NO once more to return
to Normal Operation.
7. Switching from LEL to PPM to %VOL
a. The catalytic combustible channel’s units can be changed from
% LEL (Lower Explosive Limit) to PPM (parts per million) to
% VOL (percent volume).
b. While in normal operation, press and release RANGE ▼
SHIFT until the desired unit appears.
8. Resetting and Silencing Alarms
a. If ALARM SILENCE is turned on, press and release POWER
ENTER RESET to silence an alarm. The audible alarm will
silence and the alarm LED’s will continue to pulse.
b. If the Eagle 2 alarms are set to latching, to reset an alarm,
press and release POWER ENTER RESET once the alarm
condition has cleared.
c. If the Eagle 2 alarms are set to self-resetting, the alarms will
automatically reset once the alarm condition has cleared.
9. Turning Off The Eagle 2
a. Press and hold POWER ENTER RESET. The buzzer will pulse
for about 5 seconds.
b. Release the POWER ENTER RESET button when you see
GOODBYE and the RKI logo. When these disappear, the
Eagle 2 is off.