Specifications – RKI Instruments GP-01 with self resetting alarms User Manual
Page 6

6 • Specifications
Model GP-01 Operator’s Manual
Specifications
Table 1: Model GP-01 Specifications
Target Gas
Combustible Gas, Methane (CH
4
) Calibration Standard
Detection Range
0-100% LEL (1% LEL Increments)
Detection Principle
Catalytic Combustion Method
Sampling Type
Diffusion
Response Time
T90 in 30 seconds
Indication
7-Segment Digital LCD
Gas Alarms
(Factory Settings)
•
Low Alarm: 10% LEL
•
High Alarm: 50% LEL
•
Over Range: 100% LEL
Other Alarms
Sensor Failure, Low Battery, Dead Battery, System Failure, and
Calibration Failure
Operating Temperature &
Humidity
- 20°C — 50°C, Below 90% RH (Non-condensing)
Power Supply
•
Two AAA size Alkaline Batteries Standard, Duracell MN2400 or
PC2400, Eveready Energizer E92 or EN92
•
2.4 Volt Ni-cad Battery Pack 49-1603RK Optional
Continuous Operating Time
•
Alkaline Batteries: 16 Hours Minimum at 25°C, No Alarms or
Backlighting
•
Ni-cad Battery Pack: 8 Hours Minimum at 25°C When Fully
Charged, No Alarms or Backlighting
Features
•
Manual Backlight on LCD (Automatic Lighting at Alarm)
•
Peak Hold
•
Direct Charging (Ni-cad Battery)
Standard Accessories
•
Wrist Strap
•
Rubber protective boots
Optional Accessories
•
Ni-cad Battery Pack
•
Battery Charger for Ni-cad Battery Pack
•
Calibration Adapter
•
Calibration Kit
•
Alligator Clip
•
Belt Clip
Safety/Regulatory
C US
186718
CSA certified, “C/US”, as Intrinsically Safe. Exia. Class I, Div. 1, Groups
A, B, C, & D. Temperature Code T3C.
Dimensions and Weight
1.4” (35mm) W x 4.1” (104mm) H x 0.8” (20mm) D; 2.8 oz. (80 g)