beautypg.com

RKI Instruments HS-94 User Manual

Page 8

background image

4

. Auto Zero/Span

To automatically adjust the zero point for H

2

S or CO, (or the span to 20.9% for

oxygen), take the instrument to a known fresh-air environment. With instrument in
normal operation, press and hold the AIR button. The unit will beep three times,
and then give a double beep indicating that it has adjusted to the surrounding
atmosphere.

5. Alarms

Gas Level Alarms

The 94 Series instruments have standard alarm points as listed below:

OX-94

19.5% decreasing, 23.5% increasing

CO-94

25 ppm increasing, 25 ppm TWA, 400 PPM STEL

HS-94

10 ppm increasing, 10 ppm TWA, 15 ppm STEL

When an alarm point is exceeded, the instrument sounds a pulsed tone, and the
display reading will blink. In the case of TWA or STEL alarms, the display will also
indicate which alarm level has been exceeded. The instrument reading will
automatically latch onto the highest reading (or lowest reading for low oxygen
levels). To reset, press the MODE switch once. If the gas concentration has dropped
to below the alarm point, the alarm will silence, and the unit will give the current
reading. If the unit has not dropped to below the alarm point when the MODE
switch is pressed, then the audible alarm will still sound and the unit will latch onto
the new reading present at the time the MODE switch was pressed.

Safety Self-Check

The 94 Series instruments continuously monitor themselves for proper operation. If
a malfunction occurs, a steady “trouble” tone will sound. There are two modes of
malfunction that will sound a steady

“TROUBLE” alarm. These are:

LOW BATTERY

This alarm will sound if the batteries are too low
to operate the instrument. Replace the batteries
before further use.

8

This manual is related to the following products: