Connection for central water supply – Memmert TTC Temperature test chamber additional accessories (Generation 2003) User Manual
Page 7

7
Supplement to Operating Instructions CTC/TTC
3. Connection for central water supply
With this specific variation, the climatic test chamber can be connected to a central water
supply and waste water disposal. After the waste water is introduced into the appliance, it is
deionised and demineralised by filter cartridges before it is vaporised. The connections can be
found on the rear side of the appliance.
Waste water
Ensure that the water can
run off accurately
(max. pumping height: 1 m)
Waste w
Fresh water
Limit water pressure through
a pressure controller
to max. 3 bar.
If the filter cartridges are used up, a LED will
flash on the front side of the appliance
(see picture on the right). In this case,
shut down the appliance immediately
and contact your customer service (see
operating manual page 2) to have the
filter cartridge replaced.
°C
%
rh
mb
MAX
AUTO
In case of visual alarm,
limit exceeded. Change
filter cartridge immediately.