Supplement to Operating Instructions CTC/TTC

3. Connection for central water supply

With this specific variation, the climatic test chamber can be connected to a central water

supply and waste water disposal. After the waste water is introduced into the appliance, it is

deionised and demineralised by filter cartridges before it is vaporised. The connections can be

found on the rear side of the appliance.

Waste water

Ensure that the water can

run off accurately

(max. pumping height: 1 m)

Waste w

Fresh water

Limit water pressure through

a pressure controller

to max. 3 bar.

If the filter cartridges are used up, a LED will

flash on the front side of the appliance

(see picture on the right). In this case,

shut down the appliance immediately

and contact your customer service (see

operating manual page 2) to have the

filter cartridge replaced.

°C

%

rh

mb

MAX

AUTO

In case of visual alarm,

limit exceeded. Change

filter cartridge immediately.