Supplement to Operating Instructions CTC/TTC

1.6.2 Heating up process

To accelerate the heating up process, it is recommended to set a higher chamber temperature

until the chamber load has reached the desired temperature. Afterwards, the chamber tem-

perature is also set to the desired test temperature.

Programming example:

Quickly heating up the core temperature of a test object to +20 °C. The chamber temperature

is set to +30 °C until the temperature of the test object has reached +20 °C. ASF tolerance

setting: ± 10 K.

Rampe Time

Temperature Fan

Rh

Cond

1

0:01

+30 °C

100 %

off

Spwt(T)

2

0:01

+20 °C

100 %

off

Next

3

1:00

+20 °C

100 %

off

Next

+20 °C

+30 °C

+40 °C

Temperature in the test chamber

Temperature of the test object

ASF tolerance band