Loadstar Sensors DI-100U Calibration Quick Start Guide User Manual
Page 2

Loadstar Sensors, Inc. •
www.loadstarsensors.com
WC (this will continuously output the load in lb. Change the load and the
output should change. Type
6.
To calibrate a load cell in
kg
using mVolt/V type the following commands in sequence in
Hyperterminal. Commands are not case sensitive. For example, to calibrate a 250 kg load
cell with a 3.02 mVolt/V rating:
UNITS KG
LC 250
CAL m
MVOLT 3.02 (This number will be available from the load cell
calibration/specification certificate.)
TARE (this zeros out the load)
WC (this will continuously output the load in kg. Change the load and ensure
that the output changes).
7.
To calibrate a load cell in
kg
using a known weight type the following commands in
sequence. Commands are not case sensitive. For example, to calibrate a 250 kg load cell
using a known load of 100 kg, remove all load from load cell and type:
UNITS KG
LC 250
CAL 2
2PCAL 100
<Apply Point 1 Load of 0 KG Press C when ready or Q to quit> C
<Apply Point 2 Load of 100 KG Press C when ready or Q to quit> Apply 100 kg, then type C
TARE
8.
WC (this will continuously output the load in kg. Change the load and the output should
change. Type