Labnet Spectrafuge 24D Digital Microcentrifuge User Manual
Page 6
3
Rotor
Lid
(Included)
Order
no.
C0240-RC
Rotor
Securing
Screw
(Included)
Order
no.
C0240-RSS
Strip Spin Adapter
Order
no.
C0240-SS
Tubes accepted
0.2ml tubes or 8 x 0.2ml strips
Max
Speed
13,300
rpm
Range Centrifuge radius
4.32cm to 5.33cm
Range RCF (g value)
8,549 x g to 10,539 x g
Adapter for 0.5ml tubes (pk./6)
Order
no.
C-1205
Tube
measurement
8
x
30mm
Max.
speed
13,300rpm
Centrifuging
radius
7.53cm
RCF
(g-value)
14,892
x
g
Adapter for 0.4ml tubes (pk./6)
Order
no.
C-1206
Tube
measurement
6
x
47mm
Max.
speed
13,300rpm
Centrifuging
radius
8.23cm
RCF
(g-value)
16,276
x
g
Adapter for 0.2ml tubes (pk./6)
Order
no.
C-1222
Tube
measurement
6
x
21mm
Max.
speed
13,300rpm
Centrifuging
radius
7.03cm
RCF
(g-value)
13,903
x
g
3.2 Rotor maintenance
The rotor should be cleaned thoroughly after each use. Thorough cleaning must be
performed when spinning samples containing phenol or phenol chloroform.
Periodically inspect the rotor for dents, dings, scratches, discoloration and cracks. If any
damage to the rotor is found, discontinue use of the rotor immediately and replace.
3.3 Removing and installing the angle rotor
The Spectrafuge 24D comes complete with a standard 24 place rotor installed. To remove
the rotor for cleaning, remove the rotor securing screw from the motor shaft by turning
the screw counterclockwise, using the rotor wrench. Lift the rotor directly upward in a
straight vertical motion. To replace rotor, first make sure the motor shaft and rotor
mounting hole are clean. Place the rotor on the motor shaft. Reinstall the rotor securing
screw on the motor shaft by turning it clockwise. Hold the rotor with one hand and
tighten the rotor securing screw, using the rotor wrench.