Labnet Spectrafuge 24D Digital Microcentrifuge User Manual
Page 5
2
1.5
Warranty
This centrifuge has been subjected to thorough testing and quality control. In the unlikely
event of a manufacturing fault, our one year warranty (from the date of delivery) covers
the centrifuge and the rotor. This warranty becomes invalid in the case of incorrect
operation, use of nonstandard spare parts or accessories and unauthorized modification of
the rotor or centrifuge.
Labnet reserves the right to make technical modifications.
Please see the complete limited warranty statement supplied separately.
2. Installation
2.1 Unpacking the centrifuge
Before unpacking the centrifuge, inspect the outside of the carton for any shipping
damage.
The centrifuge is delivered in a carton with protective foam cushioning. Remove the
centrifuge from the carton. Retain thecarton and cushioning until it has been established
that the centrifuge is working properly.
Inspect the centrifuge for any visible signs of shipping damage. Shipping damage is the
responsibility of the transportation carrier. Any claims for damage must be filed within
48 hours with the carrier that delivered the centrifuge.
The accessories supplied with the centrifuge should be kept with the instruction manual
near the centrifuge’s place of installation.
2.2 Required space
The centrifuge should be installed on a rigid, even surface such as a stable laboratory
bench, countertop, etc. To guarantee sufficient ventilation, ensure that the centrifuge has
at least 15cm (6 inches) of free space on all sides, including the rear.
The centrifuge should not be located in areas subject to excessive heat such as in direct
sunlight or near radiators or the exhaust of a compressor, as a buildup of heat may occur
within the chamber.
2.3
Installation
Before operating the centrifuge, check that the power source (electrical outlet on the wall)
corresponds to that on the manufacturer’s rating label, then connect the power cord to the
centrifuge and the power source.
3. Installation of rotors and rotor maintenance
3.1 Rotors and accessories
The following accessories are included or available for the Spectrafuge:
Angle rotor for 24 x 1.5ml tubes
(Included)
Tube measurement
1.5ml / 2.0ml (10 x 40mm)
Max. speed 13,300rpm
Centrifuging
radius
8.23cm
Max. RCF (g-value) 16,276 x g
Rotor
Wrench
(Included)
Order
no.
C0240-RT