Enter the sample volume. Press “Back” to go back to the file save menu. Press
“Run” to start running the file.
In the file save menu, if you press “Run” without pressing “Save”, the system will
display “The modifications will be lost
after running. Do you still want to
run?”
Press “No” to go back file save
interface.
Press “Yes” to enter the file
running menu and sample volume
input
screen.
(Note:
the
modification here cannot be
saved after the program is run.
When the file is running,
“Now
Running•••”
will continuously flash.
When the protocol is complete, “File
running is over is displayed, Press
‘Enter’ to return to the main menu.”
Press “Stop” and “Do you want
to stop running?” will be
displayed. Select ‘F1’or ‘F3’”.You
can then select either “Stop” or
“Start”.
Press “Pause” and the display
will show “Pause state now.
Press ‘F3’ to continue running.”
You can then select “Start” to
continue.
Press “View File” to view the file in the editing format.
Note: the file cannot be edited while running
Press “Skip” to stop current temperature segment and run the next segment.
Hotlid=off 2007-07-28 10:05:00
abc
User:
Labnet
i
File Name:
Num1
1
Password:
Num1
1
Password confirm:Num1
1
No Yes
s
File running interface
Hotlid=off
2007-11-1 10:05:00
100
75
50
25
0
Temp
1c 95.0
2c 55.0 3c 72.0 4 04.0
Time
1:00
02:00 01:45 00:00
File Name:Num1
User: Labnet Now Running •••
1c Temp=00.0 Time=00:00 Cycle 01 of 10
Total Time: 00h 56m Remaining Time:00h 56m
Stop
Pause Start Skip View File