Other – Labnet 611D Laboratory Incubators User Manual
Page 17
17
OTHER
Other-1 Controller on at all
times - "locked-up"
1/ Turn unit off and on to reset.
2/ If cannot change any condition on the front panel,
call Customer Service.
Other-2 Front panel displays
are all off
Check for wire damage.
Other-3 Unit or wall fuse/circuit
breaker is blown
1/ Check wall power source.
2/ Compare current draw and compare to specs on
data plate.
3/ See what other loads are on the wall circuit.
Other-4 Unit will not turn on
1/ Check wall power source.
2/ Check fuse/circuit breaker on unit or in wall.
3/ See if unit is on, e.g., fan or heater, and just
controller is off.
4/ Check all wiring connections, especially around the
on/off switch.
Other-5 Unit is smoking – Out
of box
This is not an uncommon occurrence when first
operating new units. Put unit under vent and run at
full power for one hour. – smoking is normal during
first cycle to temperature.
Other–
6
Contamination in
chamber
1/ See cleaning procedure in see section 8.0.
2/ Develop and follow standard operating procedure
for specific application; include definition of cleaning
technique and maintenance schedule.
If following the troubleshooting suggestions does not solve the problem, contact Labnet
International’s Service Department at 888 522-6381, Monday through Friday, 9:00am –
5:00pm, EST.