TROUBLESHOOTING GUIDE

PROBLEM

POSSIBLE SOLUTION

Oven will not power up or not heat

Check power cord, outlet and unit circuit breaker

Shaking platform will not operate

Check timer set to Hld or a time setting
Check that nothing is blocking shaker platform
Press shaker STOP/START button

Temperature too high

Check set point and readjust if necessary
Check calibration

Chamber temperature goes above set

Normal operation in initial heat up or if door

point and settles back to set point

opened for a long period

Temperature will not remain stable

Check that set point is at least 5C above ambient

or display shows “LO”

(or 10°C above ambient when shaker is on) which is
minimum operating point

Indicated temperature is unstable

A slight variation of +/-0.1C is normal. Larger fluctua-

tions may be ambient variations from drafts, door
opening and closing, a fan obstruction or failure or
electrical noise from RFI (motors, etc.)

Temperature is too low

If door has opened, unit may not have recovered yet.

Confirm temperature set point.
Safety thermostat set too low.

Unit will not heat above temperature

Confirm set point. Check temperature of chamber

that is below set point

with a thermometer and recalibrate if needed.

Temperature display and reference

Be sure that unit has been allowed to stabilize for

thermometer do not match

1 hour. Thermometer should be suspended in

the chamber and not touching any surface. Only
certified reference thermometers should be used.

Cannot adjust set point or calibration

Turn unit off for 5 seconds to reset. If problem

or display persists call service

persists, call for service.

Unit calibrated at one temperature

This can be a normal condition if temperatures or load

and not at another

vary widely. For best accuracy, calibrate at set point

TECHNICAL SUPPORT / SERVICE

Should your I-5311-DS shaking incubator require service, call Labnet International’s Service
Department at 732-417-0700 between the hours of 9:00am and 5:00pm EST. If service
necessitates the return of the unit to Labnet’s Service Department, a return authorization (RA)
number must first be obtained and a decontamination certificate completed. The unit must be
thoroughly decontaminated before shipping. The unit must then be sufficiently packed for
shipment. Any shipping damage on returned units will be the customer’s responsibility.