beautypg.com

JULABO TW20 Water Baths User Manual

Page 20

background image

Operating procedures

20

7.2. Switching

on / Start - Stop

Switching on:

Turn on the mains power switch (1).

The unit performs a self-test. All segments of the 4-digit LED

temperature DISPLAY and all indicator lights will illuminate.

Then the software version (example: n 11.0) appears.

Together with the display of the water bath temperature the

operating state is also displayed. (Example: 18.5 C)

The heat-up phase is indicated by the yellow heating control light

which will blink in regular intervals when the selected

temperature has been attained.

Switching off:

Turn the unit off with the mains power switch.

7.3.

Setting the temperatures

Setting can be carried out in the start/stop condition.

1.

Press one of the keys

for a short moment. The setpoint

value instead of the actual value is indicated on the display for

about 8 seconds. The value can now be changed.

2.

Change value:

Press

to set a higher value.

Press

to set a lower value.

Keep the keys depressed for the value to change fast.

3.

Press enter

to store the value.

Notice:

When the working temperature is higher than 50 °C, it might happen that due

to strong production of steam there is considerable dripping on the inside of
the lift-up Makrolon

cover. Some drops may fall directly into the material to

be tempered.

Always properly seal all sample containers.

This manual is related to the following products: