Preparations
14
6. Preparations
6.1. Installation
Caution:
The unit is not for use in explosive environment.
>80 mm
Place the waterbath in an upright position.
Keep a wall distance of minimum 80 mm.
Caution:
Potential hazards from the samples
Proper use of shaking water baths includes immersion of samples contained
in test tubes, Erlenmeyer flasks, or other containers for the purpose of
controlling their temperature.
We do not know which substances are contained within these vessels.
Many substances are:
inflammable, easily ignited or explosive
hazardous to health
environmentally unsafe
i.e.: dangerous
The user alone is responsible for the handling of these substances!
Always properly seal all sample containers.
Notice:
There is a danger of electrochemical oxidation or corrosion when using
test-tube racks or samples made of non-ferrous metal.
Avoid using these types of racks or samples.
Use only original JULABO test-tube racks.