Technical data – JULABO KRC 180 Refrigerators for chemicals User Manual
Page 16

TECHNICAL DATA
16
11. TECHNICAL DATA
Description
KRC50
KRC180
Working temperature range
-2 ... 12 °C
Temperature adjustment
analog
Temperature indication
digital, 7-Segment, green
Resolution
0,1
Temperature control accuracy
±1 K
Temperature control
two-point
Working sensor
PTC
Safety sensor
PTC
Safety installations:
Overheating protection for compressor
105 °C
Working sensor monitoring
disconnection
Working temperature adjustment
short-circuit
Safety senor monitoring
short-circuit
Monitoring of power voltages
DC
Alarm signal
optical
Alarm-Relay
Switching voltage, max.
220 V/DC; 250 V/AC
Switching current, max.
2 A
Switching power, max. (non-inductive load)
60 W / 125 VA
Contact material
Ag Pd 60/40
Volumetric capacity
50 l
180 l
Interior dimensions (WxDxH)
42 x 31 x 39 cm
52 x 40 x 70 cm
Exterior dimensions (WxDxH)
53 x 63 x 54 cm
60 x 64 x 86 cm Weight
24 kg
36 kg
Mains power connection
207...253 V / 50 Hz
Current draw (at 230 V)
0,5 A
0,5 A
Permissible ambient temperature
0 ... 35 °C