JULABO FT902 Immersion Coolers User Manual
Page 8

Operator responsibility – Safety instructions
8
2.4. Technical
specifications
Immersion cooler
FT402
Working temperature range
°C
-40 ... 30
Temperature stability
°C
±0.5
Temperature selection
digital
Temperature indication
LED
Resolution °C
0.1
Temperature control
PID1
Cooling capacity
Medium ethanol
°C
kW
+20 10 -20 -40
0.45 0.36 0.14 0.03
Cooling compressor
1-stage
Refrigerant
230 V / 50 Hz
115 V / 60 Hz
R404A
R134a
Electrical connections:
Pt100 external sensor
Pt100
Overall dimensions (WxDxH)
cm
20x30x43
Immersion probe (Lxdia.)
cm
12x5
Immersion probe, flexible (Lxdia.)
cm
-------
Connection tubing (L)
cm
120
Noise level, distance 1 m
dBA
61
Weight kg
24
Ambient temperature
°C
5 ... 35
Mains power connection
230 V/50 Hz
V/ Hz
207-253 / 50
Current input
(at 230 V)
A
4
Mains power connection
115 V/60 Hz
V/ Hz
90-127 / 60
Current input
(at 115 V)
A
4
All measurements have been carried out at: rated voltage and frequency
ambient temperature: 20 °C
Technical changes without prior notification reserved.