JULABO F34-ED Refrigerated / Heating Circulator User Manual
Page 8

Operator responsibility – Safety recommendations
8
2.4. Technical
specifications
F34-ED
Working temperature range
°C
-30 ... 100
Temperature stability
°C
±0,03
Temperature selection
digital
Temperature indication
LED
Resolution
°C
0.1
Temperature control
Heater wattage (at 230 V)
kW
2,0
Heater wattage (at 115V)
kW
1,0
Cooling capacity
Medium ethanol
°C
kW
+20 0 -20
0.45 0.32 0.14
Refrigerant
R134a
Circulating pump:
discharge, max.at 0 bar
l/min
15
pressure, max. at 0 l
bar
0.35
Overall dimensions (WxDxH)
cm
38x58x62
Bath opening (WxL)
cm
24x30
Bath depth
cm
15
Filling volume
liters
14 ... 20
Weight
kg
41
Ambient temperature
°C
5 ... 40
Mains power connection 230 V/50 Hz
V/ Hz
207-253 / 50
Current input (at 230 V)
A
12
Mains power connection 230 V/60 Hz
V/ Hz
207-253 / 60
Current input (at 230 V)
A
13
Mains power connection 115 V/60 Hz
V/ Hz
103-127 / 60
Current input (at 115 V)
A
14
Mains power connection 100 V/50-60 Hz
V/ Hz
------
Current input (at 100 V)
A
------
All measurements have been carried out at:
rated voltage and frequency
ambient temperature: 20 °C
Technical changes without prior notification reserved.