2.4. Technical

specifications

F34-ED

Working temperature range

°C

-30 ... 100

Temperature stability

°C

±0,03

Temperature selection

digital

Temperature indication

LED

Resolution

°C

0.1

Temperature control

Heater wattage (at 230 V)

kW

2,0

Heater wattage (at 115V)

kW

1,0

Cooling capacity

Medium ethanol

°C

kW

+20 0 -20

0.45 0.32 0.14

Refrigerant

R134a

Circulating pump:

discharge, max.at 0 bar

l/min

15

pressure, max. at 0 l

bar

0.35

Overall dimensions (WxDxH)

cm

38x58x62

Bath opening (WxL)

cm

24x30

Bath depth

cm

15

Filling volume

liters

14 ... 20

Weight

kg

41

Ambient temperature

°C

5 ... 40

Mains power connection 230 V/50 Hz

V/ Hz

207-253 / 50

Current input (at 230 V)

A

12

Mains power connection 230 V/60 Hz

V/ Hz

207-253 / 60

Current input (at 230 V)

A

13

Mains power connection 115 V/60 Hz

V/ Hz

103-127 / 60

Current input (at 115 V)

A

14

Mains power connection 100 V/50-60 Hz

V/ Hz

------

Current input (at 100 V)

A

------


All measurements have been carried out at:

rated voltage and frequency

ambient temperature: 20 °C

Technical changes without prior notification reserved.