beautypg.com

JULABO F34-ED Refrigerated / Heating Circulator User Manual

Page 24

background image

Cleaning / repairing the unit

24

15

A
M
P

Circulator: Mains circuit breakers (resettable) 15 A


Cooling machine: Fuse T 10.0 A, dia.5 x 20 mm

The mains fuses on the rear of the unit may easily be exchanged as
shown on the left.

Warning:

Before exchanging the fuses, turn off the mains power switch and disconnect
the power plug from the mains socket!
Only use fine fuses with a nominal value as specified.

Example:

Manufacturer Supplier

Type

Order

No.

Wickmann Wickmann G-

fuse

insert

T10,0A 5x20 mm

No. 19195











8.

Cleaning / repairing the unit

Caution:

 Always turn off the unit and disconnect the mains cable from the power

source before cleaning the unit.

 Prevent humidity from entering into the circulator.

 Electrical connections and any other work must be performed by qualified

personnel only.

To maintain the full cooling performance, clean the condenser from
time to time.

 Switch off the unit, disconnect mains power cable.
 Hold the venting grid, pull out and remove.
 Clean the ribbed condenser with a vacuum cleaner.
 Replace the venting grid.
 Switch on the unit.