beautypg.com

Environmental Express Smart Bottle User Manual

Page 10

background image

Environmental  Express,  Inc.                                                                                Smart  Bottle  Rack  Operation  and  Instruction  Manual    

 

         800.343.5319/843.881.6560  •  www.envexp.com  

 
 

9  

3.2. To change bottles or add solvent to an empty bottle first disconnect the gas pressure line from the side of

the Smart Bottle Rack. Allow sufficient time for the bottle to reach an equilibrium state with the outside
pressure. Failure to do so can result in a sudden release of pressure, which could cause injury or damage
to the surroundings. Replace or refill/empty the empty/full container and recap. Connect the gas line and
proceed with normal operations. All waste should be handled in accordance with the facility’s waste
disposal plan.

3.3. If a warning appears for low solvent/high waste it is possible to continue with the operation. The warnings

are designed to indicate when a full set of samples (6) is no longer possible. If less than a full set of
samples is being run, the warning can be cleared and the analysis may proceed.

4.0

Smart Bottle Rack Parts List

Smart Bottle Rack

G8100

DI Water and Water Miscible Waste Sensor

G8108

Hexane, Hexane Waste, and Methanol Sensor

G8109


Hexane Tubing

G8102A


DI Water Tubing

G8102B


Methanol Tubing

G8102C


Water Miscible Waste Tubing

G8102D


Hexane Waste Tubing

G8102E

Waste Carboy

G3070

Reagents

4L n-Hexane

LCCC08005

1L DI Water

LCCC08006

1L Methanol

LCCC08007