3.2. To change bottles or add solvent to an empty bottle first disconnect the gas pressure line from the side of
the Smart Bottle Rack. Allow sufficient time for the bottle to reach an equilibrium state with the outside
pressure. Failure to do so can result in a sudden release of pressure, which could cause injury or damage
to the surroundings. Replace or refill/empty the empty/full container and recap. Connect the gas line and
proceed with normal operations. All waste should be handled in accordance with the facility’s waste
disposal plan.
3.3. If a warning appears for low solvent/high waste it is possible to continue with the operation. The warnings
are designed to indicate when a full set of samples (6) is no longer possible. If less than a full set of
samples is being run, the warning can be cleared and the analysis may proceed.
4.0
Smart Bottle Rack Parts List
Smart Bottle Rack
G8100
DI Water and Water Miscible Waste Sensor
G8108
Hexane, Hexane Waste, and Methanol Sensor
G8109
Hexane Tubing
G8102A
DI Water Tubing
G8102B
Methanol Tubing
G8102C
Water Miscible Waste Tubing
G8102D
Hexane Waste Tubing
G8102E
Waste Carboy
G3070
Reagents
4L n-Hexane
LCCC08005
1L DI Water
LCCC08006
1L Methanol
LCCC08007