2

®

Fiber-Lite

WARNING

Risk of electrical shock.

Remove power plug before lamp

replacement and wait for

hot lamp to cool.

LAMP REPLACEMENT:
1. Remove all AC power from the illuminator.
2. Check the illuminator to verify that the chassis is cool before proceeding.
3. Remove the four screws holding the top cover to the chassis. Retain these screws to secure the top cover after lamp replacement.
4. Remove the top cover and set aside. The is a small wire attached to the top cover for grounding purposes. Do not disturb this wire.
5. Unplug the lamp power socket from the power supply connector. The plug is polarized to assure correct connection.
6. Lift and remove the lamp from the lamp holder by grasping the rear of the lamp adjacent to the lamp socket.
7. Remove the lamp from the socket by holding the lamp socket and gently pulling the lamp reflector.
8. Discard the lamp per instructions above. Consult

or call 1-800-895-8842.

www.lamprecycle.org

9. Insert the

CAUTION: Do not touch the interior of the lamp reflector, the lamp envelope or the lamp pins with your fingers. Touching the interior of the
lamp reflector,
of the reflector or the area adjacent to the pins.

10. Repeat steps 1- 5 in reverse order.

Lamp Socket

Ground Wire

Lamp Intensity Control:

The base Model MH100 Illuminator uses
a mechanical iris to vary the Illuminator
intensity.

Manual Iris Lever
for Intensity Control

Remove Screw

Remove Screw

Remove Screw

Remove Screw

LAMP CONTAINS MERCURY

MANAGE IN ACCORDANCE WITH DISPOSAL LAWS

REF: WWW.LAMPRECYCLE.ORG OR 1-800-895-8842