15
5. SPECIFICATION
C9501/IS C9503/IS C9505/IS C9507/IS
PSI
Range
0 to ±2
0 to ±5
0 to ±30 0 to ±100
Resolution
0.001
0.001
0.01 0.1
mBar
Range
0 to ±140
0 to ±350 0 to ±2000 0 to ±6900
Resolution 0.1 0.1 1 1
inH
2
O
Range
0 to ±55
0 to ±140
0 to ±830 0 to ±2750
Resolution 0.01 0.1 0.1 1
inHg
Range
0 to ±4 0 to ±9.999
0 to ±61 0 to ±200
Resolution
0.001
0.001
0.01 0.1
mmHg Range
0 to ±99.99
0 to ±260 0 to ±1500 0 to ±5200
Resolution 0.01 0.1 1 1
Torr
Range
0 to ±99.99
0 to ±260 0 to ±1500 0 to ±5200
Resolution 0.01 0.1 1 1
Pa
Range
0 to ±9999
Resolution
1
KPa
Range
0 to ±35
0 to ±200 0 to ±690
Resolution
0.01 0.1 0.1
mmH
2
O Range
0 to ±1400 0 to ±3500
Resolution
1 1
cmH
2
O Range
0 to ±2100 0 to ±7000
Resolution
1 1
Kgcm
–2
Range
0 to ±2.1
0 to ±7.0
Resolution
0.001
0.001
CALIBRATED
±0.2% of full scale (FS) at +23°C
ACCURACY
REPEATABILITY
±0.1% of full scale (FS) at +23°C
MEAN TEMPERATURE
<0.1% per °C
COEFFICIENT
OF READING
OPERATING
0 to 40°C, 10 to 90% RH
RANGE non
condensing
ENVIRONMENTAL
IP67
RATING
CERTIFICATION
II 1 G EEx ia IIC T4
Baseefa 03 ATEX 0079