Connection to pressure sources – Comark C9501/IS User Manual
Page 13

12
3.
CONNECTION TO PRESSURE SOURCES
The C9500 Series IS instruments are fi tted with
1
/
8
" BSP female connectors. The following checks
must be made before connecting the instrument to
a pressure source:-
Check the security of all fi ttings.
Check that the pressure source does not exceed the
ratings for the instrument as set out in Table 1.
Check that the gases to be measured will not damage
or corrode stainless steel, brass, glass, ceramic, silicon,
nickel, aluminium, gold or epoxy resin.
Note: Comark C9500 Series IS Pressure Measurement
Instruments are not suitable for absolute pressure
measurement.
Note: Pressures in excess of maximum over range
fi gure with respect to atmospheric pressure, should
not be applied to either port as this could damage the
instrument or represent a safety hazard.
TABLE 1
Instrument
Pressure Range
Maximum Over/
Under
Pressure
C9501/IS
0 to ±2 PSI/140mbar
6 PSI/400mbar
C9503/IS
0 to ±5 PSI/350mbar
15 PSI/1000mbar
C9505/IS
0 to ±30 PSI/2000mbar
90 PSI/6200mbar
C9507/IS
0 to ±100 PSI/6900mbar
200 PSI/13700mbar