3
TRANSMITTER STARTUP
The following changes must be made in the configuration file when
setting up the 40-inch and larger sensor head system.
3.1
Flow Tab
Manually input either the pipe Outer Diameter of Inner Diameter and
the Wall Thickness. There are no values for pipe sizes greater than
36 inches in the pipe Size / Schedule menu selection.
3.2
Sensor Tab
The sensor spacing values must be changed to reflect those used with
the 40-inch and larger sensor. Refer to the following Table for the
new values:
Sensor #
Standard Product
Spacing (ft)
40-inch and larger
Sensor Band
Spacing (ft)
1 0.0
0.0
2 0.2
0.4
3 0.4
0.8
4 0.6
1.2
5 0.8
1.6
6 1.0
2.0
7 1.2
2.4
8 1.4
2.8
Table 1
Sensor Spacing
P/N 21080-01 Rev 02
Page 3-1