beautypg.com

Caron 7000-75 User Manual

Page 32

background image

7000-50 & 7000-75 Series Operations Manual Rev A 1/6/2014

Page 32 of 71

Recorder Calibration:

If a calibration adjustment is required for a single pen recorder,

use the left (#1) and

right (#2) arrow push buttons on the recorder to calibrate (or move) the pen’s position

on the chart to correspond to the temperature of the solution. The arrow buttons must

be held for approximately five seconds before the pen will begin to move.

For two pen recorders, you must first select the pen that you wish to calibrate. This is

done by pressing the left (#1) arrow button to select the red pen or the right (#2) arrow

button to select the blue pen. The arrow button must be held down until the green LED

light goes out. After the green LED light goes out, follow the instructions in step #3

above.

Battery Backup:

The green LED light remains a constant green color indicating that both the battery and

the main power to the unit are good. Refer to Figure 5 for the location of the green LED

indicating light. If the AC power were to fail or the battery becomes weak, then the

green LED light will begin "flashing" indicating that either you have lost the main power

to the unit or it is time to replace the battery. Having a 9 volt DC battery back-up in

place, will allow the recorder to continue to function normally for approximately 24

hours in the event of a power failure.

This manual is related to the following products: