Safety Considerations

Read and understand this manual completely before attempting to set up or use

this instrument.

This equipment has been designed and tested to conform to CSA1010 safety standards, as

applicable to laboratory instrumentation. This applies only to the EPS-3000P-II(V) when used

as specified in the documentation, in its intended applications, with CBS Scientific approved

electrophoresis apparatus only. Usage in any other manner may not provide similar performance

or safety protection.
This equipment is provided with a 3-conductor, grounded AC line cord. The protective earth

ground is necessary for safe operation. Do not use any other AC line cord with this instrument.
The EPS-3000P-II(V) is a high voltage power supply capable of generating dangerous levels

of voltage and current during operation. Exercise caution when working around and with the

electrical connections of this equipment. Always check electrical connectors, wires, and associated

apparatus for any signs of wear or damage before using with this equipment. Be sure to use only

electrophoresis equipment that is suitably rated for the voltage and current capabilities of the EPS-

3000P-II(V) power supply.
The output of the EPS-3000P-II(V) power supply is intended for connection to electrically isolated

electrophoresis apparatus only. Use only with electrically isolated electrophoresis apparatus with

minimum isolation of 3000V. Do not connect any terminal of the EPS-3000P-II(V) output to earth

ground. This may impair the safety protection provided by the equipment, or cause equipment

damage.
The high voltage output of the EPS-3000P-II(V) power supply takes some amount of time to

decay when unloaded or lightly loaded. Wait a minimum of 60 seconds after stopping a run

before touching the power supply leads.
This equipment has a protective ground leakage current of 0.5mA at 120VAC, and 1mA

at 230VAC. While significantly below the 3.5mA limitation for laboratory instrumentation,

some applications and locations require a leakage current below 0.5mA. Check the specific

requirements of your application before using this equipment.
This equipment is for indoor use only.