3 safety – C.B.S. Scientific LPE-4003-220V User Manual
Page 6

6
LPE Instructions 5-5-14
1.3 Safety
Power to the PGGE unit is to be supplied by an external DC voltage power supply 
that must be ground isolated so that the DC voltage output floats with respect to 
ground. For any power supply used, the maximum specified operating parameters 
for the units are:
Maximum Limits
250VDC
30watts
power
125mA
current
55ºC
ambient
temperature
 
Current to the unit, provided from the external power supply, must enter the unit 
through the lid assembly, providing a safety interlock to the user. Current to the unit 
is broken when the lid is removed. 
Do not attempt to use the unit without the
safety lid, and always turn the power supply off before removing the lid, or when 
working with the unit in any way. Follow safety precautions specified by the 
power supply manufacuturer.
