beautypg.com

2 specifications, 3 safety – C.B.S. Scientific SGE-040-02 User Manual

Page 5

background image

C.B.S.  Scientific

Econo-Sub Units

5

SECTION 1
General Information
1.1 Introduction

The Econo-Sub Horizontal Gel units offer gel sizes (14 x 20cm) SGE-014-02, (20 X 20cm) SGE-
020-02, (20 x 30cm) SGE-030-02, and (20 x 40cm) SGE-040-02. The gel tray and safety cover
are UV transparent acrylic. In the Econo-Sub Horizontal gel unit, gel casting is done in the gel
tray by sealing the ends of the tray with tape. The slotted gel tray configuration allows for single or
multiple sample separations. When the gel tray is used in combination with either two or more
sample combs, designed for multi-channel pipette loading, multiple screening of large sample
numbers can be facilitated easily. The Econo-Sub Horizontal Gel units come equipped with an
integral cooling chamber for temperature control and buffer recirculation ports for maintenance of
proper pH.


1.2 Specifications

Constructions:

Buffer chamber, safety cover

Acrylic

Electrodes

Platinum wire .012" diameter

Power cords

FR Urethane rated 7500VDC,
200mA, 65°C

Tapered baffles

Acrylic

Combs

Polycarbonate, Anodized Aluminum or

Teflon

Adjustable comb backing

Acrylic

UVT Gel Tray

UV transparent acrylic

Safety Certification

EN61010-1-1993 (IEC1010-1)

SGE-014-02

SGE-020-02

SGE-030-02

SGE-040-02

Shipping weight

9 lb.

11 lb.

15 lb.

19 lb.

Overall Size
(l)x(w)x(h) cm

38 x 15 x 10

43 x 21 x 10

53 x 21 x 10

63 x 21 x 10

Gel Size cm

14 x 20

20 x 20

20 x 30

20 x 40

Recommended buffer
volume
with gel tray

1200 mls

2200 mls

2250 mls

2300 mls

Distance between
electrodes cm

33

38

48

58

Voltage limit

500 VDC

500 VDC

600 VDC

600 VDC


1.3 Safety

Power to the SGE-014-02, SGE-020-02, SGE-030-02 or SGE-040-02 units is to be supplied by
an external DC voltage power supply that must be ground isolated so that the DC voltage output
floats with respect to ground. For any power supply used, the maximum specified operating
parameters for the units are:

Maximum Limits
500-600 VDC

20watts power

90mA current
55°C ambient temperature


Current to the unit, provided from the external power supply, must enter the unit through the lid
assembly, providing a safety interlock to the user. Current to the unit is broken when the lid is
removed. Do not attempt to use the unit without the safety lid, and always turn the power
supply off before removing the lid, or when working with the unit in any way. Follow
safety precautions specified by the power supply manufacturer.