8

Mini-Fast Sub


3.3

Removing the gel

Turn the power supply off and disconnect the leads from the power supply. Remove the safety
cover from the Mini-Gel unit by placing thumbs on white posts next to red & black connectors,
then pushing down while pulling up with fingers under lid. DO NOT pull on power cords.


Gently lift the gel from the unit. Always wear gloves, eye protection and protective clothing if
buffer and/or gel contain Ethidium Bromide. Ethidium Bromide is a powerful mutagen, gloves, eye
protection and protective clothing should always be worn when handling the gel or buffer
solutions. See Material Data Safety Sheets.

View separated fragments under UV light, using proper protection for eyes and skin (see
manufacturer’s instructions).






3.4

Maximum Well/Comb Volumes

NOTE: To calculate sample well volume expressed in millimeters (mm) of height, divide
maximum volume by tooth depth.

# of wells

Tooth width

(mm)

Spacing

Between

teeth (mm)

Overall

length

(mm)

Tooth
depth

(mm)

1.0mm thickness
volume per tooth

microliters (ul)

2.0mm thickness
volume per tooth

microliters (ul)

4

20.2

2.35

88.26

5.37

108.47

216.94

8

8.88

2.28

88.26

5.37

47.69

95.38

16

3.27

2.16

88.26

5.37

17.56

35.12