5

Mini-Fast Sub




SECTION 1
General Information

1.1

Introduction

The Mini-Fast Sub, cat. #MGU-303, features a simple and safe design. Gel casting is easily
accomplished by inserting the precision fit dams and comb into the grooves provided in the body
of the unit. The gel bed dimensions are 8cm (l) x 10cm (w). This uniplanar design has a buffer
chamber only slightly longer than the gel itself, minimizing the amount of buffer required.

1.2 Specifications

Constructions

Buffer chamber, lid

Acrylic, UVT

Electrodes

Platinum wire .012" diameter

Power cords

FR Urethane, rated 7500V,
200mA, 65° C

Casting dams

PVC

Combs

PVC

Safety Certification

EN61010-1-1993 (IEC1010-1)

MGU-303

Shipping weight

3lbs

Overall Size
(l)x(w)x(h)

17cm x 14cm x 8cm

Gel Size

7.5cm x 10cm

Recommended buffer
volume mls

50ml

Distance between
electrodes

9.5cm

Voltage limit

100VDC


1.2 Safety

Power to the MGU-303 unit is to be supplied by an external DC voltage power supply that must
be ground isolated so that the DC voltage output floats with respect to ground. For any power
supply used, the maximum specified operating parameters for the units are:

Maximum Limits
100VDC voltage

50 watts power

50 mA current

55°C ambient temperature

Current to the unit, provided from the external power supply, must enter the unit through the lid

assembly, providing a safety interlock to the user. Current to the unit is broken when the lid is
removed. Do not attempt to use the unit without the safety lid. Always turn the power
supply off before removing the lid, or when working with the unit in any way. Follow safety
precautions specified by the power supply manufacturer.