PREPARATION AND POURING OF SEQUENCING GELS USING THE GEL SLIDER™

Advance preparation:

Prepare solutions, level Gel Slider, pre-coat notched glass plate, and make sure you have completed step number C-1

(page 1 of these instructions), so that side adjustment panels are correct according to the thickness of your back glass

plate (either 1/8” thick or 3/16” thick).

Step 1. Adjust width of Gel Slider by turn-

ing over and placing glass plate into

recessed area in bottom of side panels

as shown in photo 1. Adjust width

by loosening black thumb screws and

allowing side panels to increase or

decrease width of unit until plate fits

snuggly.

Step 2. Place back glass plate onto Gel Slider

up against back plate stops. Glass plate

stops are factory installed at the 42cm

position. The stops can be moved

to 22cm and 32cm position if cast-

ing shorter gels. Position spacers and

clamp to back glass plate as shown.

Place notched plate (coated side down)

at bottom of back plate, as shown in

photo2.

Step 3. Add catalyst to acrylamide solution

and fill syringe. Start casting by filling
overlapping area between plates and
beginning to slide top plate. Elapsed
time: 2:03 minutes.