Section 5
Maintenance of equipment
5.1
care and Handling
The plastic components of the EBX-700 Electrophoretic Blotting System are fabricated from polycarbonate.
Electrodes and connectors are made from pure platinum, stainless steel, and nickel plated brass. As with any
laboratory instrument, adequate care ensures consistent and reliable performance.
After each use, rinse all parts with de-ionized water. Wipe dry with a soft cloth or paper towel, or allow to air
dry. Whenever necessary, all components may be washed gently with water and a non-abrasive detergent, and
rinsed and dried as above. Never use abrasive cleaners, glass cleaning sprays or scouring pads to clean the
components, as these will damage the unit and components.
Additional precautions:
•
Do not autoclave or dry-heat sterilize the apparatus or components.
•
Do not expose the apparatus or components to phenol, acetone, benzene, halogenated hydrocarbon
solvents or alcohols.
•
Avoid prolonged exposure of the apparatus or components to UV light.
•
Do NOT treat with diethylpyrocarbonate (DEPC)-treated water for extended periods at 37
°C. A brief rinse
with DEPC-water is sufficient after a thorough wash.
5.2
Maintenance
The following inspection and maintenance procedures will help maintain the safety and reliable performance of
the EBX-700 Electrophoretic Blotting System. Replacement parts can be ordered by calling 1-800-243-4959 or
by contacting your local distributor.
• Banana plugs and power cords should be inspected regularly. If the banana plugs become loose or do not
feel friction tight replace the plugs or power cords.
• Should power cord assemblies (connectors, wire or shrouds) show any signs of wear or damage (e.g.
cracks, nicks, abrasions, or melted insulation), replace them immediately.
• The platinum wire is secured to the banana jack by compression between a stainless washer and the jack
nut. The nut/washer interface should be tight and free of corrosion.