Maintenance
New Brunswick LN2 Back-up System
English (EN)

26

8.3

Electrical connection maintenance

All electrical cables should be checked every month. Failure to do so may result in failure of the device. If
damage is found, do

not attempt to repair by yourself.

1. Turn off LN

2

Back-up System and unplug the power cord from the mains.

2. Inspect the power cord, door switch cable, solenoid cable, and temperature sensor cable for cuts or

other damage.

3. If damage is found, contact a qualifed service engineer to replace the damaged parts.

8.4

Battery maintenance

The LN

2

Back-up System runs on a rechargeable battery during a power failure.

Contact a qualified service engineer to replace the battery every 5 years or when the Low Battery Alarm
sounds despite it being plugged into the mains.

8.5

Transfer line and manifold maintenance

At least once annually, a qualified service engineer should inspect the LN

2

transfer line from the Dewar to

the injector for damage or deterioration.

WARNING! Battery

Do not attempt to replace or repair battery by yourself.

The control unit contains mains voltage.