Innova® -86 °C Freezers — Operating manual
6.4
Alarm monitoring socket
The freezers are provided with an alarm monitoring socket at the rear of the freezer and a
matching plug for external monitoring. This plug can be connected either to a central monitoring
system such as a building management system, or to a remote alarm via an auto-dialer.
Abb. 7:
Upright freezers alarm monitoring socket (except U101)
Fig. 7:
Upright freezers alarm monitoring socket (except U101)
Abb. 8:
Chest (and U101) freezers alarm monitoring socket
Fig. 8:
Chest (and U101) freezers alarm monitoring socket
The configuration of the socket is shown in (Fig. 9 on p. 26) and (Fig. 10 on p. 26), as viewed
from the rear of the freezer. Within the freezer, the socket is connected to voltage-free contacts
rated at 24 volts, 1 amp. In normal operation, with the mains/power on, pin 1 is connected to pin
2 (N/C), and in the alarm condition, with mains/power off, pin 1 is connected to pin 3.
The High Temperature Alarm output to the Remote Alarm Monitoring Socket can be programmed
to a set time delay (see Setting the alarm delay on p. 23).
1
RS-485 connector (optional)
2
Alarm monitoring socket
3
Mains/power socket
1
RS-485 connector (optional)
2
Alarm monitoring socket
3
Mains/power socket
1
2
3
1
2
3
NOTICE!
Risk of material damage
Hazardous voltages must not be connected to the remote alarm socket. Max Rating 24 V 1 A.