beautypg.com

6 ep dualfilter t.i.p.s, 7 ept.i.p.s. loretention, 8 autoclave pipet tips – Eppendorf epT.I.P.S. Version A User Manual

Page 7: Ep dualfilter t.i.p.s, Ept.i.p.s. loretention, Autoclave pipet tips, Ept.i.p.s.® — instructions for use

background image

epT.I.P.S.® — Instructions for use

7

6

ep Dualfilter T.I.P.S.

6 ep Dualfilter T.I.P.S.

During pipetting liquid flows through a small opening. A possible outcome of this are aerosols

(dispersion of solid or liquid particles in gas). Aerosols, drops and splashes could contaminate

the pipette and contaminations could be transferred to other samples.
To avoid these unintended and unnoticed contaminations ep Dualfilter T.I.P.S. pipette tips should

be used. The dualfilter is made up of hydrophobic polyethylene (PE). The first white layer defends

from splashes and drops and the second blue layer builds an aerosol barrier.
ep Dualfilter T.I.P.S. pipette tips are sterilized by electron beams and are certified

(see Certificates - verifiable safety for you on p. 9). ep Dualfilter T.I.P.S. pipette tips can not be

sterilized.

7

epT.I.P.S. LoRetention

7 epT.I.P.S. LoRetention

epT.I.P.S. LoRetention pipette tips are particularly suited to the dispensing of liquids containing

detergents. The optimized surface of the epT.I.P.S. LoRetention pipette tips counteracts the

reduced surface tension caused by detergents and also with corresponding solvents, so that the

liquid runs off without forming a film. This increases sample recovery and pipetting accuracy.
Please note:
The small amounts of residual moisture that may occur with epT.I.P.S. LoRetention pipette tips

can appear to be higher than with standard epT.I.P.S. pipette tips as the residual moisture does

not remain in the tip as a film, but collects as droplets due to the physical properties.
Due to the physical properties of the surface of epT.I.P.S. LoRetention pipette tips, the dispensing

volume is slightly lower than the nominal volume in comparison to the corresponding standard

epT.I.P.S. pipette tips when dispensing purely aqueous solutions with epT.I.P.S. LoRetention

pipette tips 0.1 – 10 μL S and 0.5 – 20 μL L. Reproducibility still has the same high performance

as standard epT.I.P.S. pipette tips.

8

Autoclave pipet tips

8 Autoclave pipet tips

1. Remove the packaging foil.
2. Autoclave the pipet tips.

Hint!

Don not use any additional desinfectants during autoclaving for box and rack.

Product

Temperature

Pressure

Time

Number of

possible

cycles

ep T.I.P.S. and ep T.I.P.S.

LoRetention

121°C

1 bar

20 min

5 times

ep T.I.P.S. Rack

121°C

1 bar

20 min

10 times

ep T.I.P.S. Box

121°C

1 bar

20 min

100 times

ep Dualfilter T.I.P.S. and ep

Dualfilter T.I.P.S.

LoRetention

Can not be autoclaved.