beautypg.com

10 troubleshooting, 1 general errors, 11 transport, storage and disposal – Eppendorf epT.I.P.S.Version B User Manual

Page 7: 1 storage, General errors, Storage

background image

Troubleshooting

epT.I.P.S.

®

English (EN)

7

10

Troubleshooting

10.1

General errors

11

Transport, storage and disposal

11.1

Storage

Symptom/
message

Cause

Remedy

Liquid is
dripping from
the tip and/or
the dispensed
volume is
incorrect.

The tip is loose or the pipette tip is
poorly fitted.

Firmly attach the pipette tips.
Use epT.I.P.S. pipette tips.

Pipetting was too quick.

The tip is withdrawn from the
liquid too quickly.

Withdraw the tip slowly from
the liquid with a time delay
(approx. 3 seconds).

Liquid could not
be aspirated.

Filter is blocked.

Carefully remove pipette tip.
Attach the pipette tip again and
retry dispensing.

Eject pipette tip. Use a new
one.

Air temperature

Relative humidity

Atmospheric
pressure

In transport
packaging

-25 °C – 55 °C

10 % – 95 %

70 kPa – 106 kPa

Without transport
packaging

-5 °C – 45 °C

10 % – 95 %

70 kPa – 106 kPa