8 autoclave pipet tips, 9 technical data, 1 ambient conditions – Eppendorf epT.I.P.S.Version B User Manual
Page 6: Autoclave pipet tips, Technical data 9.1, Ambient conditions

Autoclave pipet tips
epT.I.P.S.
®
English (EN)
6
Please note:
The small amounts of residual moisture that may occur with epT.I.P.S. LoRetention pipette
tips can appear to be higher than with standard epT.I.P.S. pipette tips as the residual
moisture does not remain in the tip as a film, but collects as droplets due to the physical
properties.
Due to the physical properties of the surface of ep T.I.P.S. LoRetention pipette tips 0.1 –
10 μL S and 0.5 – 20 μL L the dispensing volume is slightly lower than the nominal
volume when dispensing purely aqueous solutions. Reproducibility still has the same high
performance as standard epT.I.P.S. pipette tips
8
Autoclave pipet tips
1. Remove the packaging foil.
2. Autoclave the pipet tips.
9
Technical data
9.1
Ambient conditions
Do not use any additional desinfectants during autoclaving for box and rack.
Product
Temperature Pressure
Time
Number of
possible
cycles
ep T.I.P.S. and ep T.I.P.S.
LoRetention
121 °C
1 bar
20 min
5 times
ep T.I.P.S. Rack
121 °C
1 bar
20 min
10 times
ep T.I.P.S. Box
121 °C
1 bar
20 min
100 times
ep Dualfilter T.I.P.S.
ep Dualfilter T.I.P.S.
SealMax
ep Dualfilter T.I.P.S.
LoRetention
Can not be autoclaved.
Ambience
Only for use indoors.
Ambient temperature
5 °C – 40 °C
Relative humidity
10 % – 95 %, non-condensing.
Atmospheric pressure
79.5 kPa – 106 kPa