Technical data
Varipette

®

4720

English (EN)

16

1. Please note the information in the document "Decontamination certificate for product

returns".
You can find it as a PDF file on our webpage www.eppendorf.com.

2. Enter the serial number of the Varipette 4720 in the decontamination certificate.
3. With the shipment please include the completed decontamination certificate for

product returns with the Varipette 4720.

4. Send the pipette to Eppendorf AG or an Eppendorf AG service partner. The address

details of Eppendorf AG can be found on the last page of this operating manual. The list
of service partners can be found at www.eppendorf.com/worldwide.

8

Technical data

Tab. 1: Varipette 4720 and Varitip S

Tab. 2: Varipette 4720 and Varitip P

Test conditions according to ISO 8655 with original Eppendorf Varitip P or Varitip S
without valve, liquid dispensed to the inner wall of the weighing vessel.

Volume
in mL

Error

Systematic

(Incorrectness)

Random

(Imprecision)

in %

in mL

in %

in mL

2.5

± 1.0

± 0.025

± 0.2

± 0.005

5

± 0.4

± 0.02

± 0.2

± 0.01

10

± 0.3

± 0.03

± 0.2

± 0.02

Volume
in mL

Error

Systematic

(Incorrectness)

Random

(Imprecision)

in %

in mL

in %

in mL

1

± 0.6

± 0.006

± 0.2

± 0.002

5

± 0.5

± 0.025

± 0.1

± 0.005

10

± 0.3

± 0.03

± 0.1

± 0.01

Liquid

Water (demineralized according to
ISO 3696)

Reference temperature

20 °C – 27 °C, ±0.5 °C constant

Number of determinations

10