1 printing a history, 2 exporting a history as a file, 4 troubleshooting – Eppendorf Self Test Function for Mastercycler nexus User Manual
Page 8: Troubleshooting

Self Test Function for Mastercycler® pro and Mastercycler® nexus
8
Requirement
•
A user is logged on (if the
PIN
function is activated).
•
The navigation tree is displayed.
1. Open the cycler node and mark
Self-test
.
2. Click the
History
softkey.
The self-test history is displayed.
3.3.1
Printing a history
Requirement
A supported printer is connected (see Cycler operating instructions).
Click the
softkey.
3.3.2
Exporting a history as a file
1. Connect an external storage medium.
2. Click the
Export
softkey.
3. Select the PDF or TXT file format and confirm by clicking
Ok
.
The file is copied to the storage medium. The path of the copied files is displayed after the
procedure has been completed.
4
Troubleshooting
4 Troubleshooting
Symptom/
message
Cause
Remedy
Self-test USB key is
not detected.
The self-test USB key was
connected to a non-supported USB
hub.
Connect the self-test USB key
immediately.
The control panel USB connections
are not active because the control
panel was connected to a
Mastercycler ep (the symbol is
displayed).
Connect the control panel to a
Mastercycler pro immediately (see
Mastercycler pro operating
instructions).