beautypg.com

1 printing a history, 2 exporting a history as a file, 4 troubleshooting – Eppendorf Self Test Function for Mastercycler nexus User Manual

Page 8: Troubleshooting

background image

Self Test Function for Mastercycler® pro and Mastercycler® nexus

8

Requirement

A user is logged on (if the

PIN

function is activated).

The navigation tree is displayed.

1. Open the cycler node and mark

Self-test

.

2. Click the

History

softkey.

The self-test history is displayed.

3.3.1

Printing a history

Requirement

A supported printer is connected (see Cycler operating instructions).

Click the

Print

softkey.

3.3.2

Exporting a history as a file

1. Connect an external storage medium.
2. Click the

Export

softkey.

3. Select the PDF or TXT file format and confirm by clicking

Ok

.

The file is copied to the storage medium. The path of the copied files is displayed after the

procedure has been completed.

4

Troubleshooting

4 Troubleshooting

Symptom/

message

Cause

Remedy

Self-test USB key is

not detected.

The self-test USB key was

connected to a non-supported USB

hub.

Connect the self-test USB key

immediately.

The control panel USB connections

are not active because the control

panel was connected to a
Mastercycler ep (the symbol is

displayed).

Connect the control panel to a

Mastercycler pro immediately (see

Mastercycler pro operating

instructions).

This manual is related to the following products: