beautypg.com

1 operating instructions, 1 using this manual, 2 product description – Eppendorf Self Test Function for Mastercycler nexus User Manual

Page 4: 1 delivery package, 2 features, 3 operation, 1 activating the self-test function, Operating instructions 1.1 using this manual, Operation 3.1 activating the self-test function

background image

Self Test Function for Mastercycler® pro and Mastercycler® nexus

4

1

Operating instructions

1

Operating instructions

1.1 Using this manual

These instructions for use for the self-test function of the Mastercycler pro and Mastercycler

nexus is a supplement to the Thermocycler operating instructions and does not replace the

Thermocycler instructions.
Therefore, the Thermocycler operating instructions must be read before using the self-test

function for the first time. The current version of the instructions is available on the internet at

www.eppendorf.com

2

Product description

2 Product description

2.1 Delivery package

2.2 Features

The self-test function can be used with the following devices:

Mastercycler pro family of devices and the 6320 control panel (with USB ports)

Mastercycler nexus family of devices

Eliminating the need for an external measuring system, the self-test function is used to quickly

check of the following thermoblock characteristics:

Heating and cooling rates

Correct function of the temperature control circuit

Temperature homogeneity within each of the three block zones (left, center and right zone)

Temperature homogeneity across the entire block

A certificate can be output as a PDF after the self-test has been completed successfully.
The self-test USB key also activates the self-test function for all connected Mastercyclers.

3

Operation

3 Operation

3.1 Activating the self-test function

Requirement

A user is logged on (if the

PIN

function is activated).

The navigation tree is displayed.

Connect the self-test USB key to an open USB port on the Thermocycler.
The message

Please wait

appears. The self-test USB key is detected after a few seconds.

The self-test function is now unlocked.

Quantity Order No.

(International)

Order No.

(North America)

Description

1

6320 071.001

950030040

Selftest USB Key

For Mastercycler pro and Mastercycler nexus

1

6320 071.168

-

Mastercycler pro Self Test Function -

Instructions for Use