Page 6

background image

Self Test Function for Mastercycler® pro and Mastercycler® nexus

6

At block temperatures of 4 °C, 35 °C, 55 °C, 75 °C and 95 °C, the temperature

homogeneity and thermoregulation accuracy are determined after 30 s of holding time.

After 5 to 10 minutes the test result is displayed.

Passed tests are labeled

PASS

and highlighted in green.

Failed tests are labeled

FAIL

and highlighted in red.

The self-test is passed if all subtests have been passed.

Block zone

The left, center and right block zones correspond to the

temperature control loop of the block.

Heating/cooling rates

Temperature control velocity of the block during heating or

cooling.

Temperature control

Accuracy of thermoregulation.

Temperature homogeneity

Temperature homogeneity within each individual control circuit.

Overall temp. homogeneity

Temperature homogeneity across the entire block.

Hint!

Failure of a test indicates an alteration, maladjustment or malfunctioning of the block. In

this case, please contact Eppendorf Service.

Hint!

Use the Eppendorf Temperature Verification System to verifiy and adjust the

temperature control of the block (see on p. 10).

This manual is related to the following products: